Electrical conductivity is the ability of a material
to carry electrical current. In water, it is generally used as a measure of the mineral or
other ionic concentration. Conductivity is a measure of the purity of water or the
concentration of ionized chemicals in water. However, conductivity is only a quantitative
measurement: it responds to all ionic content and cannot distinguish particular conductive
materials in the presence of others.
Only ionizable materials will contribute to conductivity; materials such as sugars or oils
are not conductive. In a metal conductor, electrical current is the flow of electrons and is called electronic conductance. In water, electrical current is carried by ions since electrons do not pass through water by themselves. This is electrolytic conductance. When a voltage is applied between two inert
electrodes immersed in a solution, any ions between them will be attracted by the
electrode with the opposite charge. Ions will move between electrodes and produce a
current depending on the electrical resistance of the solution. This is the basis of
conductivity measurementan application of Ohms law. For high purity waters, it
is common to express conductivity as its reciprocal, resistivity. To prevent altering the sample by major ionic
movement and electrochemical reaction at the electrodes, alternating current is always
used for measurement. With AC the polarity changes frequently enough that ions do not move
or react significantly. Measuring systems must control the voltage, frequency and current
density to minimize errors due to electrode polarization and capacitance. Modern
instrumentation may change one or more of these variables automatically, depending on the
conductivity range being measured. Units of Measure Conductivity is the conductance of a
standardized volume of water with 1 cm2 cross-sectional area and electrodes
spaced 1 cm apart as shown in Figure 1. The same results will be obtained with various
geometries as long as the ratio of length / area is equal to 1 cm-1, the cell
constant. Units of measure are related as listed in the table below. Since most
conductivity values are quite small, unit multiplier prefixes of milli (m, 10-3)
and micro (m, 10-6) are commonly used. Units Related to Conductivity Measurement
* Most users employ units of S/cm. However,
SI conductivity units used in some parts of the world are S/m which can easily be
confused. 1 S/cm = 100 S/m. The standardized geometry of the cell
constant assures that conductivity measurement is a property of the sample and not of the
sensor. The cell constant is determined with precision by calibration in known
conductivity standard solutions. Accepted standard potassium chloride solutions are
established in an ASTM standard method. Standard reference materials are also available
from NIST.
Cell constants other than 1 cm-1
may be used as long as the measuring instrument readout is normalized accordingly. A lower
cell constant sensor is needed to enable the measuring instrument to make accurate
measurements in low conductivity (high resistivity) samples. Higher cell constants are
needed to measure in high conductivity samples. The exact requirements depend on the
measuring instrument. Most practical cells do not use the parallel
plate electrode arrangement of Figure 1. They have greater durability and allow more
convenient installation with other arrangements. For example, typical pure water sensors
for on-line measurement use concentric electrodes that maintain the spacing and geometry
for 0.01 to 0.1/cm constant. A variety of two-electrode process conductivity sensors is
illustrated in Figure 2. Conductivity of very dilute solutions can be
calculated from physical chemistry data based on Equation 1 which sums the conductivity
contribution of all ions in the solution. L = r * S (li * ci)
(1) L = conductivity r = density of water li
= equivalent ionic conductance
of ion i ci
= concentration of ion i Temperature Effects Conductivity is affected by temperature since water becomes less viscous and ions can move more easily at higher temperatures. Conventionally, conductivity measurements are referenced to 25 °C though occasionally a 20 °C reference is used. The variation with temperature is apparent in Equation 1 since li, and to a lesser degree, r are temperature dependent. The conductivity of most ions increase in conductivity by about 2.2% of their value per °C which allows for simple temperature compensation. This is suitable for most mid-range conductivity measurements. Very low and very high conductivity samples require special handling of temperature effects. As water is purified below 5 µS/cm (>0.5
MW×cm), the small amount of hydrogen (H+) and hydroxide (OH-) ions from water itself become a significant
part in the total conductivity. That is, the concentrations, ci. in Equation 1,
of H+ and OH- must be included. Water self-ionizes to a much greater
degree at higher temperatures, producing a higher concentration of ions. Those ions are
also more conductive. The temperature coefficient increases to 4 7 %/°C, depending
both on purity level and temperature, with both effects being non-linear. These effects
are evident in Figure 3. Specialized high purity temperature compensation must be employed
to achieve accurate measurements in this low conductivity range. Highly concentrated solutions also exhibit
deviant behavior. In modest concentration ranges, conductivity is roughly proportional to
concentration. At higher concentrations, Equation 1 does not apply and ionic interference
restricts the mobility of the ions. Conductivity levels off and in many cases decreases
with increased concentration as shown in Figure 4. Measurement near the peaks of these
curves is ambiguous and cannot be relied on for inference of concentration since there are
two possible values. Highly conductive solutions also have lower temperature influence,
typically less than 2 % / °C. Alternative Measurement Technologies Figure 1 and its description above refer to the conventional two-electrode sensor and measurement technique. Four-electrode conductivity measurement uses a sensor incorporating four electrodes. It is useful for highly conductive and/or dirty water samples which would foul the surfaces or plug the narrow passages of conventional high constant two-electrode sensors. Four-electrode measurement applies AC through
the sample via two outer drive electrodes as shown in Figure 5. These electrodes may
become fouled and the circuit will compensate to maintain the AC current level constant.
Two inner measuring electrodes are used to sense the voltage drop through the portion of
solution between them. The circuit makes a high impedance AC voltage measurement, drawing
negligible current and making it much less affected by additional resistance due to
fouling of the measuring electrode surfaces. Sensors for four electrode conductivity
measurement are shown in Figure 6. Inductive (also known as non-contact,
electrodeless or toroidal) conductivity measurement is made without any direct electrical
contact with the sample. The sensor consists of two parallel coils sealed within a
doughnut-shaped insulated probe as shown in Figures 7 and 8. The instrument energizes one
coil with AC. A weak AC current is induced into the surrounding sample, depending on its
conductivity. That current, in turn, induces a signal into
the measured coil which provides the measurement signal. The sample acts
like the core of a transformer. A temperature
sensor is incorporated into the probe body to enable compensation in the instrument. Because there are no electrodes in contact
with the sample, extreme fouling conditions can be tolerated. Coatings can cause errors
only if there is so much accumulation that it reduces the diameter of the hole. Care is
needed in installation of inductive sensors to allow for the specified spacing around it.
Otherwise the cross-section of sample immediately around the probe (cell constant) will be
affected and recalibration may be required. Total Dissolved Solids (TDS) TDS is sometimes inferred from conductivity
and is reported in units of parts per million. However, the relationship of conductivity
and concentration is not standardized and to be meaningful, should be specified whenever
TDS units are used. Typical conversions are based on sodium chloride (which may also be
called salinity) at approximately 0.5 ppm TDS per mS/cm. Alternatively, a natural water
mineral composition including bicarbonates would have a conversion of 0.6 0.7 ppm
TDS per mS/cm. Conversions may also be slightly
non-linear with concentration. Figures
Figure 1 cell constant = 1 cm electrode spacing divided by 1 cm2 cross-sectional area of sample
Figure 2 Conventional two-electrode conductivity sensors: 0.1 cm-1 in retractable housing, 0.1 cm-1 in flow chamber, sanitary 0.1 cm-1, 10 cm-1 insertion, 50-1 cm insertion, long 0.1 cm-1 insertion, short 0.1 cm-1 insertion.
Figure 5 Four-electrode Conductivity Measurement
Figure 6 Four-electrode Conductivity Sensors
Figure 7 Inductive Conductivity Measurement
Figure 8 Inductive Conductivity Measurement Equipment Bibliography 1. Standard Test Methods for Electrical Conductivity and Resistivity of Water, D1125, American Society for Testing and Materials, Conshohocken, PA. 2. Certificate of Analysis, Aqueous Electrolytic Conductance Standard Reference Materials 3190-3193, 3198-3199, National Institute of Standards and Technology, U. S. Department of Commerce, Gaithersburg, MD. 3. Standard Test Method for Electrical Conductivity and Resistivity of a Flowing High Purity Water Sample, D5391, American Society for Testing and Materials, W. Conshohocken, PA. 4. Weast, R. C. ed., CRC Handbook of Chemistry and Physics, 59th ed, CRC Press, West Palm Beach, FL, 1978, pp. D265D314. All figures are property of and courtesy of Mettler-Toledo/Thornton which has used and retains the rights for future use in its own presentations and publications.
|
|
|